Select Validation from the Smart Interface software main screen.
Select Run System Validation from the validation screen. Select
enclosure type, E or EN. To begin the validation procedure:
Step 1. D/A Outputs Validation
1. Select D/A Outputs. Select the output signal you wish to
monitor for this test, 4-20 mA, 0-5 VDC or 0-10 VDC
(optional). Click Continue.
2. Connect the flow meter as shown on the screen. For the Zero
Condition Test, enter the current output value as displayed on
the digital multimeter. Click Continue.
3. For the Span Condition Test, enter the current output value as
displayed on the digital multimeter. Click OK to continue. The
Smart Interface software returns to the validation screen and in-
dicates the D/A Outputs pass/fail status.
Step 2. Electronics Validation
1. Select Electronics to begin the next validation step. (The user
full scale and/or the k-factor are automatically adjusted to the re-
quired factory defaults for this test.)
2. Access the Smart electronics device and locate jumper J1. Move
this jumper from the RUN position to the CAL position.
3. Connect the flow meter as shown on the screen.
4. To validate the 0% condition, turn the CAL pot until the digital
multimeter indicates the displayed BV value. Enter the measured
value in the Indicated Value selector. Click OK to continue.
5. Repeat Step 4 to validate the 25%, 50%, 75% and 100% con-
ditions.
6. When complete, replace jumper J1 to the RUN position. The
Smart Interface software returns to the validation screen and in-
dicates the Electronics pass/fail status. (Also, the user full scale
and K-factor are automatically returned to their original settings.)
Warning!
Do not open the flow
meter enclosure in a
hazardous area.