Using the Validation Screens
When selecting Validation from the main screen, the software dis-
plays the current calibration data for the attached flow meter includ-
ing the bridge voltages (raw signal from the sensor) used for cali-
bration, the flow meter’s calibration parameters, the totalizer settings
and the sensor temperature and resistance values.
You may use the Validation screen to:
Read Curve. This routine allows you to view and alter (password
protected) the calibration data stored within the flow meter’s Smart
electronics.
Run System Validation. This routine guides you through the entire
validation process and allows you to save and print the data.
Select Reset Totalizer to reset the totalizer to zero (the password is
not required). Make sure the desired range is selected before reset-
ting the flow totalizer.
Figure 4. Validation Screen
Press CTRL-H to toggle
the Smart Interface help
screen. Help screens
are available when the
mouse is positioned
over a command field or
dialog box.
Select Reset Totalizer to
reset the current range to
a zero value.
Select Run System Validation to
perform validation procedures.
Select Read Curve to view
calibration data.