Instruction manual
Model 5100-99-IT 4-20 mA Module 
4.4 CONFIGURE SET-POINTS 
The transmitter set-points menu is used to initially set-up the alarm set points, relay actions, range, 4-20 mA 
action and RS-485/Sentry address and baud rates. 
•  Alarm Set-points: Once the Set-up menu is selected, press [E] to activate the Alarm Set-point 
screen. Use the [▲] or [▼] keys to select Low Alarm or High Alarm menu. Key [▲] will adjust the 
setpoint upwards and Key [▼] will adjust the value downwards.  Once it reaches the desired 
setpoint, Key [E] will accept it and ACK will appear. 
•  Alarm Relay Set-up: Once the Set-up menu is selected, press key [▼] once and press [E] to 
activate the Relay Set-up menu. Use the [▲] or [▼] keys to select High Alarm or Low Alarm relay 
menu and press [E]. Use the [▲] or [▼] keys to select the correct alarm relay action for the 
application, Latch, Sentry or Non-Latch. Selecting “Sentry” enables the Sentry controller to make all 
alarm action decisions. * indicates the current selection. 
•  Range: Once the Set-up menu is selected, press key [▼] twice and press [E] to activate the Range 
Set-up menu. Use the [▲] or [▼] keys to select Range menu and press [E]. When “Range” is 
selected menu provides any choices of ranges available for the gas type selected. Use the [▲] or 
[▼] keys to select the desired range. If the “User” range is selected, use the [▲] or [▼] keys to 
adjust the high end of the range desired. See specifications (page 25) for optional ranges. 
•  4-20mA: Once the Set-up menu is selected, press key [▼] three times and press [E] to activate the 
4-20mA Set-up menu. Use the [▲] or [▼] keys to select Calib or CalibOut menu and press [E].   
The “Calib” section of the menu allows the user to calibrate the 4 mA and 20 mA outputs. To 
calibrate the 4 mA and 20 mA outputs it is necessary to have an amp meter connected to the 5100-
XX-IT and upon selecting the 4 mA output calibration then the [▲] or [▼] keys can be used to adjust 
the 4 mA reading on the amp meter until it reads 4 mA. Similar steps can then be performed for the 
20 mA output. The CalibOut section allows the user to select the 4-20 mA output action desired 
during calibration. * indicates the current selected value. Available selections include: 
  Track – the 4-20mA value tracks the calibration gas exposed to the transmitter 
  Zero – the 4-20mA value is held at 0mA during calibration 
  C1.50mA – the 4-20mA value is held at 1.50mA during calibration 
  C4.00mA – the 4-20mA value is held at 4.0mA during calibration. 
  TblOut – the TblOut menu allos user to select 2.5 mA or A user selected value for 
Trouble mA signal. 
  DigInput – the DigInput menu allows user to select supervised or non-supervised input 
  WarmMin – the WarmMin menu allows user to select warm up time in minutes. 
•  RS-485 - Once the Set-up menu is selected, press key [▼] four times and press [E] to activate the 
RS-485/Sentry Set-up menu. Use the [▲] or [▼] keys to select “Mode” or “Calib” menu and press 
[E]. The “Mode” menu allows the user to select units including %LEL, % Vol, PPM % Full Scale, or 
NA. The “Calib” section of the menu allows the user to calibrate the 4 mA and 20 mA outputs and 
inputs. Note that the 5100-99-IT has a rotary switch on the faceplate and it is used to select 
addresses 1-15. When connected to Sentry the user can select 1-8 and when using Modbus RS-
485 the user can select addresses 1-15. For Modbus addresses above 15, set the rotary switch to 
0 and then use the “Address” menu to select any address between 16 and 254. The Baud rate 
menu allows the user to select a baud rate of 38400, 19200, 9600, 4800 or 2400. * indicates current 
selection. Sentry default is 9600. 
Page: 17 