Data Sheet for Product

2
Smart Infrastructure
CE2N5324en
2021-03-
19
Use
The UH50.. heat and cooling energy meter physically acquires energy consumption. The
amount of heat or cooling energy is acquired separately and can be read out either on site or
from a remote location. The device consists of a flow measuring section and processor unit.
The processor unit can be equipped with different sensors, different types of power supply
and communication modules. The UH50.. is used to measure the consumption of heat in dis-
trict heating systems and blocks of flats. It is also suited to measure the consumption of cool-
ing energy (cooling energy only or in combination with thermal energy), or volume flow in hy-
dronic systems.
Restrictions
Do not use the UH50.. to measure potable water.
Also, the product is not suited for use in systems operating on water-glycol mixtures.
Functions
Ultrasonic measuring principle
The flow is acquired based on the non-wear ultrasonic measuring principle, which requires
no moving parts.
The quantity of energy supplied to the consumer by the medium during a certain period of
time is proportional to the temperature differential of flow and return and the volume flow
passing through the meter.
The water volume is measured in the measuring tube by ultrasonic pulses which are trans-
mitted in the direction of flow and against the direction of flow. Downstream, the time re-
quired by the pulses to travel from the transmitter to the receiver is reduced, upstream it is
increased. The water volume is calculated based on the difference in runtime.
The flow and return temperatures are determined using platinum resistors.
The water volume as well as the temperature difference between the warmer and cooler side
is multiplied using a heat coefficient and the product is integrated. The result, which is the
consumed heating or cooling energy quantity, is registered and displayed in the physical
units kWh/MWh or MJ/GJ; the volume in m
3
.
Processor unit
The electronics unit includes a built-in service unit. A common electronic unit applies to all
flow variables.
Optical communication interface
The meter is supplied with an optical close-range interface. The meter is equipped with an
optical communication interface which facilitates readout and configuration on site with the
help of the optical read head WZR-OP-USP and matching UltraAssist software.
Self diagnostics
The meter continuously performs self-diagnostics, allowing it to detect a number of mounting
or device errors and to display them.