Instruction manual

SITRANS TW
114 A5E00054075-05
Functions / Operation by HART
Figure 17 Principle circuit diagram simulation
4.2.18.1 General information
As long as the simulation is switched on, the transmitter does not react to sensor
input signals.
No simulation is possible in difference or average circuit.
If thermocouples are simulated, a fixed reference point temperature of 0°C is
used for the simulation.
Simulation of the electronics temperature has no influence on the analog output.
It can only be observed through the HART communication interface.
4.2.18.2 Simulation of measuring input
Simulation as a fixed value
Under consideration of the physical unit, fixed simulation values can be parameter-
ized for both simulation lines (primary measured value and electronics temperature).
The analog output value is set according to the value defaulted for the primary meas-
ured value.
Simulation with a periodic ramp function
In addition to the adjustable fixed values a periodically recurring ramp function can
also be parameterized for both simulation paths. A settable start and end value
determines the respective limits between which the simulation values move with a
Measuring channel 1
Measuring channel 2
T electronics
Measuring
Simulation
values
I constant
Values primary
Measured value
values
T electronics
Set value
Ramp
Measured value calculation
primary
measured value
secondary
measured value
tertiary
measured value
T electronics
e.g.