Instruction manual
SITRANS TW
100 A5E00054075-05
Functions / Operation by HART
• Certificates and licenses
− Information whether or not the transmitter may be operated in intrinsically safe
mode (this specification can only be read).
• Other parameterizable functions are e.g.:
− Slave pointer functions
− Simulation of measuring input, electronics temperature and analog output
− Message relay settings
− Line resistance measurement
− Sensor trim function with selectable trim range within the measuring range
limits
− Trimming the analog output
− Self-test function of hardware and firmware
− Factory reset: Resetting of the operating data to the ex-factory state.
The operating data are stored in a non-volatile memory (EEPROM).