Dose Information
31
For spiral scan protocols, the indicated mAs is the
effective mAs per image. The correlation between tube
current mA and effective mAs of spiral scans on a Mul-
tislice CT scanner is given by the following formula:
where Slice collimation refers to the collimation of one
detector row, and nrow is the number of used detector
rows.
Pitch Factor =
Feed/Rot
nrow x Slice collimation
mA =
effective mAs
x Pitch Factor
R o t T i m e
Effective mAs = mA x RotTime/Pitch Factor