Scanner User Manual

Table Of Contents
syngo Fly Through
261
Standing Point
The cone rotates around the viewing point.
Viewing Point
The complete cone moves.
Clipping Point
The cone rotates around the standing point.
Viewing Angle
Is changed by moving one of the sides of the
angle.
Viewing Distance
The distance from the standing to the viewing
point.
Viewing Depth
Moving the Front/Back Clip Plane changes the
viewing depth.
Front Clip Plane
To remove foreground-obscuring tissue. The posi-
tion of the front clip plane will always be between
standing and viewing point.
•Back Clip Plane
To remove tissue at the back of the volume. The
position of the back clip plane will always be
behind viewing point.
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 261 Friday, April 8, 2005 9:55 AM