Scanner User Manual

Table Of Contents
174
Vascular
BodyAngioFast
Indications:
Spiral mode for longer coverage and larger vessels.
A range of 20 cm will be covered in 18 sec.
BodyAngio 2
nd
reconstruction
kV 130
Effective mAs 55
Rotation time 0.8 sec
Slice
collimation
2.5 mm
Slice width 5.0 mm 3.0 mm
Feed/Rotation 10.0 mm
Pitch Factor 2.0
Increment 5.0 mm 2.0 mm
Kernel B31s B31s
CTDIVol 5.93 mGy
Effective dose Male: 2.23 mSv
Female: 2.77 mSv
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 174 Friday, April 8, 2005 9:55 AM