Scanner User Manual

Table Of Contents
Thorax
107
Hints
If you want to reconstruct thin slices every 15 or
20 mm instead of 10 mm as predefined, simply
change the Feed/Scan before loading the mode.
With studies of interstitial changes in the lungs, con-
trast medium is not necessary.
ThorHRSeq
kV 130
Effective mAs 100
Rotation time 1.0 sec
Slice collimation 1.0 mm
Slice width 1.0 mm
Feed/Scan 10.0 mm
Kernel U90s
CTDIVol
2.16 mGy
Effective dose Male: 1.02 mSv
Female: 1.33 mSv