Scanner User Manual

Table Of Contents
Thorax
103
Hints
You could repeat the same protocol by simply click-
ing the chronicle with the right mouse button for
repeat“, e.g., when both non-contrast and contrast
studies are required.
For lung cancer evaluation, this protocol can be com-
bined with the protocol “Neck Routine”.
ThorRoutine 2
nd
reconstruction
kV 130
Effective mAs 60
Rotation time 0.8/1.0 sec
Slice
collimation
4.0 mm
Slice width 5.0 mm 5.0 mm
Feed/Rotation 14.4 mm
Pitch Factor 1.8
Increment 5.0 mm 5.0 mm
Kernel B41s B70s
CTDIVol
6.47 mGy
Effective dose Male:
3.28
mSv
Female:
4.23
mSv
Contrast medium IV injection
Start delay 25 – 30 sec.
Flow rate 2.5 ml/sec.
Total amount 80 ml
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 103 Friday, April 8, 2005 9:55 AM