Thorax
107
Hints
If you want to reconstruct thin slices every 15 or
20 mm instead of 10 mm as predefined, simply
change the Feed/Scan before loading the mode.
With studies of interstitial changes in the lungs, con-
trast medium is not necessary.
ThorHRSeq
kV 130
Effective mAs 100
Rotation time 1.0 sec
Slice collimation 1.0 mm
Slice width 1.0 mm
Feed/Scan 10.0 mm
Kernel U90s
CTDIVol
2.16 mGy
Effective dose Male: 1.02 mSv
Female: 1.33 mSv