User's Manual
2-37
1010NFM-3J
Section 2
2.6 THE FLOW/TOTAL UNITS MENU
The Flow/Total Units menu is available after selecting a meter type and measurement channel. Use the
Flow/Total Units menu to select volumetric flow units and an associated time base for the energy and
flow rate and total outputs. After making your selections, a view-only menu cell shows the resultant
scaling. Another menu cell lets you adjust the output resolution by selecting a display range.
This system provides three totalizer display modes:
POSFLOW Positive (forward)/flow total only
NEGFLOW Negative (reverse)/flow total only
NETFLOW The net of the Positive and Negative/flow (default)
Invoking the NOTOT command disables the Totalizer. To-
talization will not resume until you repeat the Fn and Nu-
meric 2 key sequence. When you activate NOTOT, an N
precedes the TOTAL symbol (i.e., [NTOTAL]) on the LCD
Screen.
*Use the <F1> key as the “Lead-in command” for 4-Path Totalizer operations.
The LAPTOT command freezes the Totalizer screen
display. However, the flow computer will continue to up-
date its internal registers. The flow computer will show the
current total when you repeat the F1-Numeric 3 key se-
quence. When you activate LAPTOT, an L precedes the
TOTAL symbol (i.e., [LTOTAL]) on the LCD Screen.
Clears the Batch Sample totalizer register. See Batch/
Sample Tot.
CLEAR
TOTALIZER CONTROLS (the ‘n’ in <Fn> = channel number)*
(Makeup Latch)
(Totalizer snapshot)
(Totalizer Freeze)
Resetting the Totalizer registers clears all total data ac-
cumulated during operation.
NOTE: In Dual-Beam mode, the Totalizer operates only on
the virtual system channel (Ch 3). Therefore in this case,
the CLEAR trigger would be <F3> <1>.
NOTOT
(also clears overflow)
F1
F2
F3
F4
6
F1
F2
F3
F4
4
F1
F2
F3
F4
3
F1
F2
F3
F4
2
F1
F2
F3
F4
1
Clears the Makeup Latch. See paragraph 2.7 The Data
Span/Set/CAL Menu and subparagraph 2.7.1 Span Data.
CLEAR
(Batch/Tot Register)
CLEAR
LAPTOT