User's Manual
2-22
1010NFM-3JSection 2
2.4.1 HOW TO SELECT A TRANSDUCER MODEL
Use the [Transducer Model] menu cell to define the type of transducer for use with your application.
This allows the flow computer to adjust its transmit/receive functions accordingly. You can choose
from either the [1011 Universal] or [1011H High Precision] transducer lists, or from a list of [991 Univer-
sal] models.
You should consider using high precision transducers for extremely critical applications. We usually
recommend these transducers for custody transfer, leak detection or nuclear power applications. The
system default is [1011 Universal]. If this suits the application, bypass this menu cell by pressing the
<Down Arrow>.
Pick/Install Xdcr Transducer Model 991 Universal
1011 Universal
1011 Hi Precision
Transducer Size
A1A1 0A0A A1HA1H
A2 0 A2H
B1 1 A3H
B2 2A B1H
B3 2 B2H
C1 3A C1H
C2 3 C2H
C3 4A D1H
D1 4 D2H
D2 5A D4H
D3 5
E1
E2
E3
Xdcr Mount Mode Direct
Reflect
Spacing Offset Minimum
Nominal
Maximum
Number Index 4 (generated)
Spacing Method Spacer Bar [P/N] auto P/N generation
Ltn Value (in) 2.00 (generated)
Install Completed? No [Yes] [Yes] indicates successful install
Install
Install Transfer
Empty Pipe Set MTYmatic
Set Empty
Actual MTY
Zero Flow Adjust AutoZero Reflect mount only
ZeroClr
Reversamatic
Actual Zero
ZeroMatic
1011 Universal 991 Universal 1011H High Precision
PICK/INSTALL XDCR MENU STRUCTURE