Use this menu to select
a class of liquids and
optionally edit the
following liquid
parameters:
•
Estimated Vs
•
Viscosity
•
Density
(Specific Gravity)
Siemens 2 Channel [1] Channel 1
Select Liquid Class from Liquid Table
UniMass Table Create/Edit Table
Temperature Range -40F to 250F
Pipe Configuration Fully Developed
Anomaly Diams 10
Liquid Class Water 20C/68F
Application Data
The UniMass Table is
used to associate the
metered liquid’s current
temperature and sonic
velocity with its specific
gravity and viscosity.
Use this menu cell to
select an upstream
pipe compensation
correction table.
Use this menu cell
to specify the
temperature range
of the liquid.
Use this menu cell to enter
the number of pipe
diameters between the
upstream configuration and
the transducer installation.
APPLICATION DATA MENU STRUCTURE
Application Data Liquid Class Select Liquid 20% Ethylene Glyclol
50% Ethylene Glyclol
27% CaCl Brine
Water 13C/55F
Water 20C/68F
Water 50C/122F
Water 75C/167F
Water 100C/212F
Water 125C/257F
Water 150C/302F
Water 175C/347F
Water 200C/392F
Water 225C/437F
Water 250C/482F
Acetic Acid
Alcohol
Bromine
Carbon Tet
Chlorine
Diesel Fuel
Gasoline
Glycerine
Kerosene
MEK
Oil (SAE 20)
Sea Water
Toluene
Trichloroethyl
Other
(continued)
The Temperature Range menu cell allows you to specify the expected temperature range at the trans-
ducer mounting location. The default setting: [-40
0
F to 250
0
F], matches the standard 991 and 1011
series of transducers but not the 1011H. For higher pipe temperatures, selecting the proper tempera-
ture range allows the flow computer to recommend the appropriate transducers.