User's Manual
4-15
1010NFM-3J
Section 4
To store the data press . This moves the cursor to Network ID.
4.8.5 NETWORK ID
The [Network ID] menu cell stores an identification number to facilitate host computer polling when you
use this system in a network environment. The Network ID number can be any value other than 0
(zero). The default setting, zero, disables the network function. Note that entering a non-zero Network
ID number suspends all routine Datalogger activity regardless of any selected options.
NOTE: If you are using the flowmeter for a “stand-alone” application, you must keep the zero
Network ID number.
To assign a Network ID number
To enable numeric entry press .
Use the numeric keys to type the Network ID number.
To store Network ID press . This moves the cursor to RTS Key Time.
4.8.6 RTS KEY TIME
During a serial transmission session, you can select how long the flow computer holds its request-to-send line
high until it receives a clear-to-send signal. If the RTS time expires, the 1010 will return its RTS line low and abort
the transmission attempt. Data entry units are seconds. The default RTS key time is 0.2 seconds.
NOTE: The RTS Key Time function is only provided when the meter’s network ID number is
non-zero.
To set the RTS Key Time:
To access the RTS Key Time option list press
To move the asterisk to the required RTS time press
To store the data press
4.9 BACKLIGHT
This menu cell allows you to specify how long you want to maintain the backlighting for the LCD
Graphic Display. Select [On] for continuous illumination. Press the <Right Arrow> to access the option
list and then scroll to desired time delay (before turn-off).
Press to store your setting.
4.10 SYSTEM INFO
This menu provides general information about the flowmeter.