User's Manual
2-70
1010NFM-3JSection 2
Siemens Dual Path 4SS10G
Access Path/Channel Diagnostic Data
Channel /Path Setup
Pipe Data
Application Data
Pick/Install Xdcr
Operation Adjust
Flow/Total Units
Data Span/Set/Cal
Stripchart Setup
Datalogger Setup
I/O Data Control
Diagnostic Data
Dual Path Flow
Siemens Dual Path 4SS10G
Transducer Setup Data
Path Select 1
Path Enable Yes
Flow Rate
Application Info
Liquid Data
Site Seup Data
Test Facilities
Print Site Setup No
Date Site Created: 04.09.0314.52.23
Diagnostic Data
“Manual” Adjustment Procedure
z In the [Site Setup Data] Menu, press the <Down Arrow> and scroll to the [HF] menu cell. Press
the <Right Arrow> and a pop-up [Manual] prompt will appear as shown below. Note: Press the
<Up/Down Arrow> to select [Automatic], if desired.
z Use the numerical keys to input the desired correction value. Press <ENT> to input value.
fx (drive) 29
N (burst length) 5
Ltn in 5.316
Vf max CU Adjustment
Vs max m/s >Manual
Vs min m/s 1355.00
Samples/Cycle 29
HF >-0.000
Siemens Dual Path 4SS10G
Site Setup Data
fx (drive) 29
N (burst length) 5
Ltn in 5.316
Vf max CU >Manual
Vs max m/s = -0.120
Vs min m/s 1355.00
Samples/Cycle 29
HF >-0.000
Siemens Dual Path 4SS10G
Site Setup Data
z In the [Diagnostic Data] Menu, highlight [Path Select] and select the desired transducer path.
Press <ENT> to select path.
z Press the <Down Arrow> and scroll to the [Site Setup Data] menu cell. Press the <Right
Arrow> to select it.
z The new correction value will appear next to the [HF] menu cell as shown below.