Data Sheet for Product
13
Siemens
A6V11160938_en
--
_c
Smart Infrastructure 2019-04-10
Error info
Error info 0-10 V output Modbus LED indication LCD
Replace
sensor
module
Present 0 V (2
s) and 10 V (2 s)
one by one in
turn
Value of register 209
changes from 0 to 1
Red permanently on
Communicati
on error
Present 0 V (5
s) and 10 V (5 s)
one by one in
turn
Value of register 209
changes from 0 to 2
Red flashing (0.5 s on /
0.5 s off)
Warning for
possible
inaccurate
measurement
Present the
measured value
Value of register 209
remains 0 without
change
Red / yellow flashing
(0.5 s red / 0.5 s
yellow)
Diagrams
Wiring
G, G0: Operating voltage AC 24 V (SELV) / DC 13.5…35 V
U1: Signal output DC 0...10 V for PM2.5 with selected range in
μg/m3
U2: Signal output DC 0...10 V for PM10 with selected range in
μg/m3
Gu: Signal ground for U1 & U2
+,-, Ref: Modbus signal
Dimensions