Scanner User Manual

Children
559
* The conversion factor for a 7-year-old child, and a
scan range of 200 mm was used.
Emotion 16 AbdRoutine
kV 110
Effective mAs/
Quality ref. mAs
60
Rotation time 0.6 sec.
Acquisition 16 x 1.2 mm
Slice collimation 1.2 mm
Slice width 5.0 mm
Feed/Rotation 28.8 mm
Pitch Factor 1.50
Increment 5.0 mm
Kernel B41s
CTDI
Vol
4.32 mGy
Effective dose Male: 2.78 mSv*
Female: 3.38 mSv*
Emotion 6 AbdRoutine
kV 110
Effective mAs/
Quality ref. mAs
60
Rotation time 0.8/0.6 sec.
Acquisition 6 x 2.0 mm
Slice collimation 2.0 mm
Slice width 5.0 mm
Feed/Rotation 18.0 mm
Pitch Factor 1.50
Increment 5.0 mm
Kernel B41s
CTDI
Vol
4.26 mGy
Effective dose Male: 7.95 mSv*
Female: 12.17 mSv*