Scanner User Manual
Radiation Therapy
481
Emotion 16 Pelvis 2
nd
recon.
kV 130
Effective mAs/
Quality ref. mAs
150
Rotation time 1.0 sec.
Acquisition 16 x 1.2 mm
Slice collimation 1.2 mm
Slice width 3.0 mm 3.0 mm
Feed/Rotation 28.8 mm
Pitch Factor 1.50
Increment 3.0 mm 3.0 mm
Kernel B41s B60s
CTDI
Vol
16.80 mGy
Effective dose Male: 6.68 mSv
Female: 11.42 mSv
Emotion 6 Pelvis 2
nd
recon.
kV 130
Effective mAs/
Quality ref. mAs
150
Rotation time 1.0 sec.
Acquisition 6 x 2.0 mm
Slice collimation 2.0 mm
Slice width 3.0 mm 3.0 mm
Feed/Rotation 18.0 mm
Pitch Factor 1.50
Increment 3.0 mm 3.0 mm
Kernel B41s B60s
CTDI
Vol
16.35 mGy
Effective dose Male: 6.41 mSv
Female: 10.95 mSv