Scanner User Manual

242
Thorax
For the 3
rd
and 4
th
reconstruction a non-square matrix
3D recon job is predefined.
Emotion 6 Thor
Combi
2
nd
recon.
3
rd
recon.
4
th
recon.
kV 130
Effective
mAs/
Quality ref.
mAs
70
Rotation
time
0.8 sec.
Acquisition 6 x 1.0 mm
Slice
collimation
1.0 mm
Slice width 5.0 mm 5.0 mm 5.0 mm 5.0 mm
Feed/
Rotation
10.8 mm
Pitch
Factor
1.80
Increment 5.0
mm 5.0 mm 5.0 mm 5.0 mm
Kernel B41s B70s B41s B70s
CTDI
Vol
8.89 mGy
Effective
dose
Male: 4.52 mSv
Female: 5.85 mSv