Scanner User Manual
Thorax
233
Emotion 16 ThorRoutine 2
nd
recon.
kV 130
Effective mAs/
Quality ref. mAs
70
Rotation Time 0.6 sec.
Acquisition 16 x 1.2 mm
Slice collimation 1.2 mm
Slice width 5.0 mm 5.0 mm
Feed/Rotation 15.40 mm
Pitch Factor 0.80
Increment 5.0 mm 5.0 mm
Kernel B41s B70s
CTDI
Vol
7.84 mGy
Effective dose Male: 4.36 mSv
Female: 5.55 mSv
Emotion 6 ThorRoutine 2
nd
recon.
kV 130
Effective mAs/
Quality ref. mAs
70
Rotation Time 0.8/0.6 sec.
Acquisition 6 x 2.0 mm
Slice collimation 2.0 mm
Slice width 5.0 mm 5.0 mm
Feed/Rotation 10.2 mm
Pitch Factor 0.85
Increment 5.0 mm 5.0 mm
Kernel B41s B70s
CTDI
Vol
7.63 mGy
Effective dose Male: 3.87 mSv
Female: 5.02 mSv