Contrast Medium
123
The dynamics of the contrast enhancement is deter-
mined by:
Patient cardiac output
•Injection rate
Total volume of contrast medium injected
Concentration of the contrast medium
Type of injection – uni-phasic or bi-phasic
Patient pathology
Aortic time-enhancement curves after i.v. contrast
injection (computer simulation*).
All curves are based on the same patient parameters
(male, 60-year-old, 75 kg).
* Radiology 1998; 207:647 – 655