Service manual

Compressor Compact Disassembling and assembling
E382 EH81E 061 01 03 02 Siemens-Elema AB 37
Compressed air reservoir
To remove the air reservoir
Refit the transport safety locks as
described in section the Transport safety
locks.
Turn the Compressor Compact upside
down. Note – The compressor
must not be
turned upside down if the compressor
casing is removed.
Remove the compressed air outlet
nipple (1).
Compressors equipped with standby
function: remove the compressed air inlet
nipple (2).
Disconnect the moisture relief valve
tube (3).
Remove the three clamps (4).
Lift off the complete air reservoir (5).
To refit the air reservoir
The procedure for refitting the air reservoir
is the reverse of the removal procedure
previously described.
COMP033X
4
51
2
3