Specifications

2/59
Siemens FI 01 · 2015 US Edition
Temperature Measurement
Transmitter for field mounting/field indicator
SITRANS TF - Transmitter, two-wire system and SITRANS TF - Field indicator for 4 to 20 mA
2
Overview
Our field devices for heavy industrial use
•HART, Universal
4 to 20 mA, universal
Field indicator for 4 to 20 mA signals
The temperature transmitter SITRANS TF works where others
feel uncomfortable.
Benefits
Universal use
- as transmitter for resistance thermometer, thermocouple ele-
ment, or mV signal
- as field indicator for any 4 to 20 mA signals
Local sensing of measured values over digital display
Rugged two-chamber enclosure in die-cast aluminium or
stainless steel
Degree of protection IP67
Test terminals for direct read-out of the output signal without
breaking the current loop
Can be mounted elsewhere if the measuring point
- is hard to access,
- is subject to high temperatures,
- is subject to vibrations from the system,
- or if you want to avoid long neck tubes and/or protective
tubes.
Can be mounted directly on American-design sensors
Wide range of approvals for use in potentially explosive atmo-
spheres. "Intrinsically safe, non-sparking and flameproof" type
of protections, for Europe and USA.
SIL2 (with Order Code C20), SIL2/3 (with C23)
Application
SITRANS TF can be used everywhere where temperatures need
to be measured under particularly adverse conditions, or where
a convenient local display is ideal. Which is why users from all
industries have opted for this field device. The rugged enclosure
protects the electronics. The stainless steel model is almost
completely resistant to sea water and other aggressive ele-
ments. The inner workings offer high measuring accuracy, uni-
versal input and a wide range of diagnostic options.
Function
Configuration
The communication capability over the HART protocol V 5.9 of
the SITRANS TF with an integrated SITRANS TH300 permits pa-
rameterization using a PC or HART communicator (hand-held
communicator). The SIMATIC PDM makes it easy.
Parameterization is carried out using a PC for SITRANS TF with
the integrated and programmable SITRANS TK. Available for
this purpose are a special modem and the software tool
SIPROM T.
Mode of operation
Mode of operation of SITRANS TF as temperature transmitter
The sensor signal, whether resistance thermometer, thermocou-
ple element or or mV signal, is amplified and linearized. Sen-
sor and output side are electrically isolated. An internal cold
junction is integrated for measurements with thermocouple ele-
ments.
The device outputs a temperature-linear direct current of 4 to
20 mA. As well as the analog transmission of measured values
from 4 to 20 mA, the HART version also supports digital commu-
nication for online diagnostics, measured value transmission
and configuration.
SITRANS TF automatically detects when a sensor should be in-
terrupted or is indicating a short-circuit. The practical test termi-
nals allow direct measurement of 4 to 20 mA signals over an am-
meter without interrupting the output current loop.
Mode of operation of SITRANS TF as field indicator
Any 4 to 20 mA signal can be applied to the generous terminal
block. As well as a range of predefined measurement units, the
adjustable indicator also supports the input of customized units.
This means that any 4 to 20 mA signal can be represented as
any type of unit, e.g. pressure, flow rate, filling level or tempera-
ture.
Mode of operation: SITRANS TF with integrated transmitter and digital
display
1 Analog-to-digital converter
2 Electrical isolation
3 Microprocessor
4 Digital-to-analog converter
5 Power supply
6 PC/laptop
7 HART modem
8 Digital display (option)
9 Test connector
Sensor
Power supply
Load
0000
00°C
TC
RTD
A
A
D
D
µP
SITRANS TF
SITRANS TH300
1
34
2
5
6
7
8
9
© Siemens AG 2015