User`s manual

20 Installation 2000591-001
Site Wiring Requirements, Continued
Cable Type Vendors & P/N Installation Notes
Twisted Pair Copper
Wire used in SAA
2100 Installations
SAA P/N C00128
Belden P/N 9462
Not Recommended
for New Installa-
tions
Nominal capaci-
tance 155 pF/m (50
pF/ft.)
Max distance from
analyzer to ANG or
between hubs 225
meters (750 ft)
All other Copper Wire Not Recommended
Customer Respon-
sibility for testing
and conformity re-
quirements
The following information pertains when using copper wire
See Page 22, CE Installations.
Use approved Conduit wiring method. Conduit may contain multiple ca-
bles. Adhere to all Federal and local electrical code requirements.
Cable Type: Refer to Table 2-1.
Use Cable Gland wiring method. Each cable gland allows entry of one
cable.
Cable Type: Recommend Belden 9182 or armored Belden 9182 or
equivalent. The proper cable gland must be used to connect the armor or
shield to the enclosure. Other cables with individually shielded pairs can
be used, but decreased distance is allowed (typically ~800-1000 feet for
22 to 26 AWG low capacitance communication cable ~ 50 to 500 is al-
lowed.
Table 2-1. Recommended
DataNET Cable,
Continued
Copper Wire Site
Requirements
CE Installations
NEC Div 2 or Non-
Hazardous Rated Areas
Cenelec Zone 1 or Zone 2
Rated Areas