56 Clinitek Advantus Operator’s Guide
Troubleshooting
5. Remove the fixed platform by pulling the entire assembly towards you.
Figure 20 Remove the Fixed Platform
6. Remove the holddown plate from the fixed platform. Refer to Performing
the Daily Cleaning‚ page 37.
7. Remove the jammed strip.
8. Reinstall the holddown plate. Refer to Performing the Daily Cleaning‚
page 37.
9. Reinstall the fixed platform. Refer to Performing the Daily Cleaning‚
page 37.
10. Reinstall the push bar. Refer to Performing the Daily Cleaning‚ page 37.
11. Turn analyzer power on.
12. Rerun the specimen(s) without results.