42 Clinitek Advantus Operator’s Guide
Maintenance
18. Reinstall the fixed platform:
a. Align the 2 grooves on the bottom of the fixed platform with the arms
extending forward from the analyzer.
The flanges on the sides of the holddown plate align just outside the
read area cover. The top edge of the platform aligns just under the
cover.
b. Gently push the platform in as far as possible.
Push past the ridge to correctly position the platform.
19. Reinstall the push bar:
a. Hold the push bar at the indented end.
b. With this end slightly upward, insert the peg on the other end of the
bar into the hole in the pusher mechanism.
c. Lower the push bar into place.
20. Place a new liner into the waste bin.
21. Clean the display screen, with a soft, nonabrasive cloth dampened with a
mild glass cleaner.
22. Turn analyzer power on.
CAUTION
Do not force the platform. Ensure that the moving table is correctly
positioned before you attempt to reinstall the fixed platform. If you
force the platform, you may damage the moving table or fixed
platform.
CAUTION
Do not use bleach to clean the display. Do not spray or pour the
glass cleaner directly onto the screen. Do not use laboratory wipes,
such as Kimwipes, because they may scratch the screen.