Clinitek Advantus Operator’s Guide 41
Maintenance
12. Dry each piece with a paper towel or soft cloth.
Use care when drying around the pins on the moving table.
13. Allow the calibration bars on the platform to air dry.
14. After cleaning, inspect the calibration bars for scratches, marks, or
discoloration.
If you cannot clean the bars, discard the current platform and replace it
with a new one.
15. Disinfect the parts, if required. Refer to Performing a Decontamination‚
page 43.
NOTE: Do not disinfect the liner. Discard it into an appropriate container and
use a new liner.
16. Reinstall the moving table:
a. Hold the table with the small rectangular tab facing to the back.
b. Align the 2 grooves on the bottom of the table with the edges of the
platform on which the table rests.
c. Gently push the table in until you hear the tab latch into the hold
position.
d. Check that the table is secure.
17. Reinstall the holddown plate:
a. Position the holddown plate with the arrow side facing up and the
arrow pointing to the back.
b. Place the pin on the front of the holddown plate into the hole at the
front of the fixed platform.
c. Align the tab at the back of the holddown plate with the slot at the back
of the platform.
d. Snap the holddown plate into place.
e. Ensure that the white calibration bars are visible.