40 Clinitek Advantus Operator’s Guide
Maintenance
8. Remove the holddown plate from the fixed platform:
a. Press upwards on the tab at the back of the plate.
b. Pull the other end from its retaining hole.
NOTE: You must remove the holddown plate for proper cleaning.
Figure 11 Remove the Holddown Plate
9. Clean the push bar, the platform, the holddown plate, and the table with
warm water and mild detergent.
10. If the holddown plate or push bar is extremely dirty, soak it in warm water
and mild detergent to loosen the dried residue.
11. Rinse each piece thoroughly.
1
Tab
CAUTION
Do not use any type of solvent to clean the analyzer. Harsh
chemicals can damage the platform components.
CAUTION
When cleaning the platform, avoid wiping across the 2 white
calibration bars. Use a cotton-tipped swab, wetted with plain water,
to clean the bars. Cleaning solution can damage the calibration
bars.