User's Manual
Table Of Contents
- Safety Notes
- FCC Conformity
- CE Electromagnetic Compatibility (EMC) Conformity
- Industry Canada
- The Manual
- Technical Support
- Abbreviations and Identifications
- SITRANS LR560 Overview
- Specifications
- Installation
- Wiring
- Local Operation
- Operating via SIMATIC PDM
- Functions in SIMATIC PDM
- SIMATIC PDM Version
- Quick Start Wizard via SIMATIC PDM
- Changing parameter settings using SIMATIC PDM
- Operating via FDT (Field Device Tool)
- Parameter Reference
- 1. Quick Start
- 2. Setup
- 2.1. Identification
- 2.2. Device
- 2.3. Sensor
- 2.4. Signal Processing
- 2.4.1. Near Range
- 2.4.2. Far Range
- 2.4.3. Minimum Sensor Value
- 2.4.4. Maximum Sensor Value
- 2.4.5. Echo select
- 2.4.6. Sampling
- 2.4.7. Echo Quality
- 2.4.8. TVT setup
- 2.4.9. TVT shaper
- 2.4.9.1. Breakpoint 1-9
- 2.4.9.2. Breakpoint 10-18
- 2.4.9.3. Breakpoint 19-27
- 2.4.9.4. Breakpoint 28-36
- 2.4.9.5. Breakpoint 37-45
- 2.4.9.6. Breakpoint 46-54
- 2.4.9.7. Breakpoint 55-63
- 2.4.9.8. Breakpoint 64-72
- 2.4.9.9. Breakpoint 73-81
- 2.4.9.10. Breakpoint 82-90
- 2.4.9.11. Breakpoint 91-99
- 2.4.9.12. Breakpoint 100-108
- 2.4.9.13. Breakpoint 109-117
- 2.4.9.14. Breakpoint 118-120
- 2.5. AIFB1
- 2.6. AIFB2
- 2.7. Measured Values
- 2.8. Filtering
- 3. Diagnostics
- 4. Service
- 5. Communication
- 6. Security
- 7. Language
- Appendix A: Alphabetical Parameter List
- Appendix B: Troubleshooting
- Appendix C: Maintenance
- Appendix D: Technical Reference
- Principles of Operation
- Echo Processing
- Measurement Response
- Damping
- Loss of Echo (LOE)
- Temperature derating curve
- Appendix E: PROFIBUS PA Profile Structure
- Appendix F: Communications via PROFIBUS PA
- Appendix G: Firmware Revision History
- Glossary
- Index
- LCD menu structure
Page 68 SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS 7ML19985LT01
mmmmm
Parameters
2.3.5.3. Sensor Offset
A constant offset (negative or positive) that can be added to sensor value
1)
to compensate if the sensor reference point has shifted. (For example, this
could result from adding a thicker gasket or reducing the standoff/nozzle
height.) The units are defined in Unit (2.3.1.).
2.3.5.4. Low Level Point
The level when the material is at Low Calibration Point. The unit is defined
in Level Unit (2.3.2.).
2.3.5.5. High Level Point
The level when the material is at High Calibration Point. The unit is defined
in Level Unit (2.3.2.).
2.3.5.6. Level Offset
A constant offset that can be added to Level. The unit is defined in Level
Unit (2.3.2.).
2.3.6. Rate
2.3.6.1. Response Rate
Sets the reaction speed of the device to measurement changes.
.
Use a setting just faster than the maximum filling or emptying rate
(whichever is greater).
1)
The distance from sensor reference point to target (see
Minimum Sensor Value 2.4.3.
on page 70
for an illustration.
Values
Range: -100 to 100. Default: 0 m
Related parameters
Unit (2.3.1.)
Values
Range: -999999 to 999999.
Default: 0%
Values
Range: -999999 to 999999.
Default: 100%
Values
Range: -999999 to 999999.
Default: 0%
Note: Default settings in the parameter tables are indicated with an asterisk
(*) unless explicitly stated.
Notes:
• Changing Response Rate resets Fill Rate/Min (2.3.6.2.), Empty rate/
Min (2.3.6.3.), and Filter Time Constant (2.5.8.1.).
• Selecting SLOW changes setting for Average amount (2.8.3.) to 0.9.
Response Rate
(2.3.6.1.)
Fill Rate/Min (2.3.6.2.)/
Empty rate/Min (2.3.6.3.)
Filter Time Constant
(2.5.8.1.)
Slow
0.1 m/min (0.32 ft/min)
10 s
* Medium 1.0 m/min (3.28 ft/min) 10 s
Fast 10.0 m/min (32.8 ft/min) 0 s