Quick Start: local
Requesting an Echo Profile
In PROGRAM mode, navigate to: Level Meter >
Diagnostics (3.) > Echo Profile (3.1.).
Press RIGHT arrow to request a profile.
•Use UP or DOWN arrow to scroll to an icon. When an icon is highlighted
that feature becomes active.
To move a cross-hair, press RIGHT arrow to increase the value, LEFT arrow
to decrease.
To Zoom into an area, position the intersection of the cross-hairs at the center of
that area, select Zoom, and press RIGHT arrow . Press LEFT arrow to Zoom
out.
To update the profile, select Measure and press RIGHT arrow .
To return to the previous menu, select Exit then press RIGHT arrow .
Device Address
The unique address of the device on the network (also called PROFIBUS address).
a) In PROGRAM mode, navigate to: Level Meter > Communication (5.) > Device
Address (5.1.).
b) Press RIGHT arrow twice to open parameter view and enable Edit mode.
c) If required, key in a new value and press RIGHT arrow to accept it. The LCD
displays the new value.
d) Press Mode to return to Measurement mode.
Note: See
Master Reset
on page 55 to reset Device Address to 126.
Values
0 - 126. Default: 126
algorithm: tF (trueFirst)
confidence
distance from flange face to target
TVT
echo
measure
exit
distance from Low Calibration Point to vertical cross-hair
zoom
pan left/right - selected
pan up/down
Echo amplitude (in dB)