Page 36 SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS 7ML19985LT01
mmmmm
Quick Start: local
Units
Low Calibration Point
High Calibration Point
Wizard Complete
To transfer Quick Start values to the device and return to Program menu, press
DOWN arrow (Finish). Then press LEFT arrow to return to Measurement mode.
Operation Description
LEVEL (1) *
Distance from Low Calibration Point to material surface.
SPACE (2)
Distance from High Calibration Point to material surface.
DISTANCE (3)
Distance from Sensor Reference Point to material surface.
Options
BACK, CANCEL, FINISH (Display returns to 1.1 Quick Start Wizard menu when
Quick Start is successfully completed.)
Sensor measurement units.
Options
M, CM, MM, FT, IN
Default: M
Operation
High Calibration
Point
Low Calibration Point
Level
Space
Distance
sensor reference point
(flange face: see
Dimensions
on page 12)
Distance from Sensor Reference Point to
Low Calibration Point: usually process
empty level.
Values
Range: 0.0000 to 100.000 m
Distance from Sensor Reference Point to
High Calibration Point: usually process full
level.
Values
Range: 0.0000 to 100.000 m