User's Manual

Table Of Contents
Page 140 SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS 7ML19985LT01
mmmmm
F: Communications PA
S:25
Internal error. Reset power. If fault persists,
contact your local Siemens
representative.
3
1
S: 27
Internal device failure caused by
an External RAM memory error.
Repair required: contact your
local Siemens representative.
3
S:28
Internal device failure caused
by a RAM memory error.
Repair required: contact your
local Siemens representative.
4
S:29
EEPROM damaged.
Repair required: contact your
local Siemens representative.
5
S: 30
EEPROM corrupt. Reset power. If fault persists,
contact your local Siemens
representative.
6
S:31
Flash error.
Repair required: contact your
local Siemens representative.
7
S: 32
IDENT number conflict.
Ensure value of the Ident
number selector is correct for
the network configuration. If it is
correct, the device needs to be
re-parameterized by the PLC.
4
0
S:33
Factory calibration for the
internal temperature sensor
has been lost.
Repair required: contact your
local Siemens representative.
1
S:34
Factory calibration for the
device has been lost.
Repair required: contact your
local Siemens representative.
2
S:39
Transducer temperature sensor
failure.
Repair required: contact your
local Siemens representative.
7
S:40
Transducer temperature too
high.
Relocate device and/or lower
process temperature enough to
cool device. Inspect for heat-
related damage and contact your
local Siemens representative if
repair is required.
5 0
Acyclic Extended Diagnostics /General Fault Codes (cont’d)
LCD
display
Meaning (cont’d) Corrective Action
Byte
(cont’d)
Bit