User's Manual
Table Of Contents
- Safety Notes
- FCC Conformity
- CE Electromagnetic Compatibility (EMC) Conformity
- Industry Canada
- The Manual
- Technical Support
- Abbreviations and Identifications
- SITRANS LR560 Overview
- Specifications
- Installation
- Wiring
- Local Operation
- Operating via SIMATIC PDM
- Functions in SIMATIC PDM
- SIMATIC PDM Version
- Quick Start Wizard via SIMATIC PDM
- Changing parameter settings using SIMATIC PDM
- Operating via FDT (Field Device Tool)
- Parameter Reference
- 1. Quick Start
- 2. Setup
- 2.1. Identification
- 2.2. Device
- 2.3. Sensor
- 2.4. Signal Processing
- 2.4.1. Near Range
- 2.4.2. Far Range
- 2.4.3. Minimum Sensor Value
- 2.4.4. Maximum Sensor Value
- 2.4.5. Echo select
- 2.4.6. Sampling
- 2.4.7. Echo Quality
- 2.4.8. TVT setup
- 2.4.9. TVT shaper
- 2.4.9.1. Breakpoint 1-9
- 2.4.9.2. Breakpoint 10-18
- 2.4.9.3. Breakpoint 19-27
- 2.4.9.4. Breakpoint 28-36
- 2.4.9.5. Breakpoint 37-45
- 2.4.9.6. Breakpoint 46-54
- 2.4.9.7. Breakpoint 55-63
- 2.4.9.8. Breakpoint 64-72
- 2.4.9.9. Breakpoint 73-81
- 2.4.9.10. Breakpoint 82-90
- 2.4.9.11. Breakpoint 91-99
- 2.4.9.12. Breakpoint 100-108
- 2.4.9.13. Breakpoint 109-117
- 2.4.9.14. Breakpoint 118-120
- 2.5. AIFB1
- 2.6. AIFB2
- 2.7. Measured Values
- 2.8. Filtering
- 3. Diagnostics
- 4. Service
- 5. Communication
- 6. Security
- 7. Language
- Appendix A: Alphabetical Parameter List
- Appendix B: Troubleshooting
- Appendix C: Maintenance
- Appendix D: Technical Reference
- Principles of Operation
- Echo Processing
- Measurement Response
- Damping
- Loss of Echo (LOE)
- Temperature derating curve
- Appendix E: PROFIBUS PA Profile Structure
- Appendix F: Communications via PROFIBUS PA
- Appendix G: Firmware Revision History
- Glossary
- Index
- LCD menu structure
7ML19985LT01 SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS Page 139
mmmmm
F: Communications PA
S:8
Service interval as defined in
Maintenance Required Limit
has expired.
Perform service.
1
0
S:9
Service interval as defined in
Maintenance Demanded Limit
has expired.
Perform service.
1
S:10
Input parameters High
Calibration Point and Low
Calibration Point are the same.
• Check calibration settings of
device.
• Ensure settings for High Cali-
bration Point and Low Calibra-
tion Point are different.
3
S:11
Internal temperature sensor
failure.
Repair required. Contact your
local Siemens representative.
4
S:12 *
Internal temperature of the
device has exceeded
specifications: it is operating
outside its temperature range.
• Relocate device and/or lower
process temperature enough
to cool device.
• Inspect for heat-related dam-
age and contact your local Sie-
mens representative if repair is
required.
• Fault code will persist until a
manual reset is performed
using PDM or the LCD inter-
face.
5
S:14
Upper and lower input values
(Process Value Scale) for AIFB1
are the same.
• Check configuration for AIFB1.
• Ensure that Upper Value and
Lower Value (Process Value
Scale) are not the same.
6
S:15
Upper and lower input values
(Process Value Scale) for AIFB2
are the same.
• Check configuration for AIFB2.
• Ensure that Upper Value and
Lower Value (Process Value
Scale) are not the same.
7
S: 17
Calibration interval as defined in
Maintenance Required Limit
has expired.
Perform calibration.
2
1
S: 18
Calibration interval as defined in
Maintenance Demanded Limit
has expired.
Perform calibration. 2
Acyclic Extended Diagnostics /General Fault Codes (cont’d)
LCD
display
Meaning (cont’d) Corrective Action
Byte
(cont’d)
Bit