Page 126 SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS 7ML19985LT01
mmmmm
E: Profile structure
2) Level Calibration is a linear transfer function that converts a sensor value to a level
value.
1)2)
3) The Transducer Block provides three possible outputs
Primary Value (PV): Level (level units)
3)
Secondary Value 1 (SV1): Level (level units)
Secondary Value 2 (SV2): Distance (sensor units)
1)
Sensor Offset (2.3.5.3.) is a constant offset (negative or positive) that can be added to
sensor value to compensate if the sensor has been changed.
2)
Level Offset (default O) can compensate for specific vessel configurations.
3)
LR560 PA does not support Volume.
Level
High Level Point
Low Level Point
Calibration
Level Value
Low
Calibration
Point
HIgh
Calibration
Point
Sensor
Value
High Level Point
(default: 100%)
sensor
reference
point
(flange
face)
sensor
value
1)
Low Level Point
(default: 0%)
Level/PV
Level Offset
2)
(if used)
Level /SV1
Low
Calibration
Point
High
Calibration
Point
Distance/SV2
SITRANS LR560