7ML19985LT01 SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS Page 113
mmmmm
B: Troubleshooting
Display shows
S: 0 LOE
antenna malfunction:
temperature too high
physical damage
check temperature in Maximum
Value (3.4.2.)
•relocate
Reading does not
change, but the level
does
SITRANS LR560 processing
wrong echo, i.e. vessel
wall, or structural member
re-locate SITRANS LR560
check nozzle for internal burrs or
welds
rotate instrument 90°
•use Auto False Echo Suppression
(2.4.8.1.) if necessary: see
Auto False
Echo Suppression (2.4.8.1.)
on page
118
Measurement is
consistently off by a
constant amount
setting for Low Calibra-
tion Pt. not correct
setting for Sensor Offset
not correct
check distance from sensor refer-
ence point to Low Calibration Pt.
(2.3.5.1.)
•check Sensor Offset (2.3.5.3.)
Screen blank
power error check nameplate rating against volt-
age supply
check power wiring or source
Reading erratic echo confidence weak refer to Confidence (2.4.7.1.)
•useAuto False Echo Suppression
(2.4.8.1.) and Auto False Echo Sup-
pression Range (2.4.8.2.)
material filling re-locate SITRANS LR560
Reading response
slow
Fill Rate/Min (2.3.6.2.)
setting is incorrect
increase measurement response if
possible
Reads correctly but
occasionally reads
high when vessel is
not full
detecting close range
echo
build up near top of ves-
sel or nozzle
nozzle problem
Use the air purge feature to clean the
antenna
•use Auto False Echo Suppression
(2.4.8.1.) and Auto False Echo Sup-
pression Range (2.4.8.2.)
Level reading lower
than material level
material is within Near
Range zone
multiple echoes pro-
cessed
decrease Near Range (2.4.1.): mini-
mum value depends on antenna type
raise SITRANS LR560
ensure Algorithm (2.4.5.1.) is set to tF
(trueFirst echo)
Operation Troubleshooting (cont’d)
Symptom Cause Action