7ML19985LT01 SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS Page 109
mmmmm
B: Troubleshooting
S: 3 Device is nearing its lifetime limit
according to the value set in
Reminder 1 (Required) (4.7.5.).
Replacement is recommended.
S: 4 Device is nearing its lifetime limit
according to the value set in
Reminder 2 (Demanded) (4.7.6.).
Replacement is recommended.
S: 6 Sensor is nearing its lifetime limit
according to the value set in
Reminder 1 (Required) (4.8.5.).
Replacement is recommended.
S: 7 Sensor is nearing its lifetime limit
according to the value set in
Reminder 2 (Demanded) (4.8.6.).
Replacement is recommended.
S: 8 Service interval as defined in
Reminder 1 (Required) (4.9.5.) has
expired.
Perform service.
S: 9 Service interval as defined in
Reminder 2 (Demanded) (4.9.6.) has
expired.
Perform service.
S:10
Input parameters Low Calibration Pt.
(2.3.5.1.) and High Calibration Pt.
(2.3.5.2.) are the same.
Check calibration settings of
device.
Ensure settings for High Calibra-
tion Point and Low Calibration
Point are different.
S: 11 Internal temperature sensor failure. Repair required: contact your local
Siemens representative.
General Fault Codes (cont’d)
Code /
Icon
Meaning Corrective Action