7ML19985LT01 SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS Page 99
mmmmm
Parameters
4.10.4. Reminder Activation
Allows you to enable a maintenance reminder
.
a) First set the values in Reminder 1 (Required) (4.10.5.)/Reminder 2
(Demanded) (4.10.6.).
b) Select the desired Reminder Activation option.
4.10.5. Reminder 1 (Required)
If Time until Next Calibration (4.10.3.) is equal to or less than this value, the
device generates a Maintenance Required reminder.
a) Modify values as required.
b) Set Reminder Activation (4.10.4.) to the desired option.
4.10.6. Reminder 2 (Demanded)
If Time until Next Calibration (4.10.3.) is equal to or less than this value, the
device generates a Maintenance Demanded reminder.
a) Modify values as required.
b) Set Reminder Activation (4.10.4.) to the desired option.
4.10.7. Maintenance Status
Indicates which level of maintenance reminder is active.
In SIMATIC PDM, open the menu View – Device Diagnostics, select Device
Status, click on the Maintenance tab, and check the Calibration Schedule
Status window.
Note: To modify this parameter via SIMATIC PDM it must be accessed via the
pull-down menu Device – Maintenance.
Values
*Timer OFF
ON
NO LIMITS
ON REMinder 1 (Maintenance Required) checked
ON
REMinders 1 - 2 checked
ON
REMinder 2 (Maintenance Demanded) checked
Values
Range: 0 to 20 years
Default: 0.164 years
Values
Range:
0 to 20 years
Default: 0.019 years