User's Manual

Table Of Contents
7ML19985LY01 SITRANS LR560 (FF) – INSTRUCTION MANUAL Page 93
mmmmm
Parameters
2.3.6.2. Fill Rate/Min
Defines the maximum rate at which the reported sensor value
1)
is allowed
to increase. Allows you to adjust the SITRANS LR560 response to increases
in the actual material level. Fill Rate is automatically updated whenever
Response Rate (2.3.6.1.) is altered.
Enter a value slightly greater than the maximum vessel-filling rate, in units
per minute.
2.3.6.3. Empty rate/Min
Defines the maximum rate at which the reported sensor value
1)
is allowed
to decrease. Adjusts the SITRANS LR560 response to decreases in the
actual material level. Empty Rate is automatically updated whenever
Response Rate (2.3.6.1.) is altered.
Enter a value slightly greater than the vessel’s maximum emptying rate, in
units per minute.
1)
The value produced by the echo processing which represents the distance from sen-
sor reference point to the target (see Minimum Sensor Value (2.4.3.) on page 95 for an
illustration).
Options
Range: 0 to 999 999 m / min.
Response Rate (2.3.6.1.)
Fill Rate
Slow 0.1 m/min (0.32 ft/min)
*
Medium 1.0 m/min (3.28 ft/min)
Fast 10.0 m/min (32.8 ft/min)
Related
parameters
Level Unit (2.3.2.)
Options
Range: 0 to 999 999 m / min.
Response Rate (2.3.6.1.)
Empty Rate
Slow 0.1 m/min (0.32 ft/min)
*
Medium 1.0 m/min (3.28 ft/min)
Fast 10.0 m/min (32.8 ft/min)
Related
parameters
Level Unit (2.3.2.)