Page 38 SITRANS LR560 (FF) – OPERATING INSTRUCTIONS 7ML19985LY01
mmmmm
Quick Start: local
Units
Low Calibration Point
High Calibration Point
Wizard Complete
To transfer Quick Start values to the device and return to Program menu, press
DOWN arrow (Finish). Then press LEFT arrow to return to Measurement mode.
Options
BACK, CANCEL, FINISH (Display returns to 1.1 Quick Start Wizard menu when
Quick Start is successfully completed.)
Sensor measurement units.
Options
M, CM, MM, FT, IN
Default: M
Distance from Sensor Reference Point to
Low Calibration Point: usually process
empty level
Values
Range: 0.000 to 40.000 m or
0.000 to 100.000 m
Default: 100.000 m
sensor reference point
(flange face: see
Dimensions
on page
12)
High Calibration Point
Low Calibration Point
Distance from Sensor Reference Point to
High Calibration Point: usually process full
level.
Values
Range: 0.000 to 40.000 m or
0.000 to 100.000 m
Default: 0.000 m