Page 24 SITRANS LR560 (FF) – INSTRUCTION MANUAL 7ML19985LY01
Wiring
5) Connect the wires to the terminals as shown (SITRANS LR560 FF is not polarity-
sensitive).
1)
2)
6) Ground the instrument according to local regulations.
7) Tighten the gland to form a good seal.
8) Replace optional display. .
9) After programming and device configuration, secure the locking screw and replace
device lid.
1)
May be shipped with the device.
2)
The instrument shield connection is internally connected to the external ground
lug.
Notes:
Foundation Fieldbus (HI) must be terminated at both extreme ends of the cable for
it to work properly.
Please refer to the
Foundation Fieldbus System Engineering Guidelines (AG-181)
Revision 2.0
, available from www.fieldbus.org, for information on installing FF (H1)
devices.
cable gland
1)
(or NPT cable entry)
cable shield
instrument
shield
connection
2)