User's Manual
Table Of Contents
- Safety Notes
- FCC Conformity
- CE Electromagnetic Compatibility (EMC) Conformity
- Industry Canada
- The Manual
- Technical Support
- SITRANS LR560 Overview
- Specifications
- Installation
- Wiring
- Local Operation
- Operating via AMS Device Manager
- Functions in AMS Device Manager
- Programming via AMS Device Manager
- Adding a new device
- Startup
- Configuring a new device
- Changing Block Modes
- Configure/Setup Parameters
- Transducer Block Parameters
- Level Transducer Block Parameters
- Operation (LTB)
- Simulation (Input)
- Setup (LTB)
- General
- Calibration
- Rate
- Signal Processing (LTB)
- Range
- Echo Select
- Sampling
- Echo Quality
- Filtering
- TVT Setup
- TVT Shaper 1
- TVT Shaper 2
- TVT Shaper 3
- Echo Profile Parameters (view only)
- Maintenance & Diagnostics (LTB)
- Sensor Lifetime
- Service Schedule
- Communication (LTB)
- Communication:
- Configure/Setup (Liquid Crystal Display Block-LCD)
- Configure/Setup (Diagnostic Transducer Block-DIAG)
- Configure/Setup (Resource Block - RESOURCE)
- Device Diagnostics (Level Transducer Block - LTB)
- Device Diagnostics (Level Control Device Block - LCD)
- Device Diagnostics (Diagnostic Transducer Block - DIAG)
- Device Diagnostics (Resource Block - RESOURCE)
- Password Protection
- AMS Menu Structure
- Parameter Reference
- 1. Quick Start
- 2. Setup
- 2.1. Identification
- 2.2. Device
- 2.3. Sensor
- 2.4. Signal Processing
- 2.4.1. Near Range
- 2.4.2. Far Range
- 2.4.3. Minimum Sensor Value
- 2.4.4. Maximum Sensor Value
- 2.4.5. Echo Select
- 2.4.6. Sampling
- 2.4.7. Echo Quality
- 2.4.8. TVT (Auto False Echo Suppression) Setup
- 2.4.9. TVT Shaper
- 2.4.9.1. Shaper 1-9
- 2.4.9.2. Shaper 10-18
- 2.4.9.3. Shaper 19-27
- 2.4.9.4. Shaper 28-36
- 2.4.9.5. Shaper 37-45
- 2.4.9.6. Shaper 46-54
- 2.4.9.7. Shaper 55-63
- 2.4.9.8. Shaper 64-72
- 2.4.9.9. Shaper 73-81
- 2.4.9.10. Shaper 82-90
- 2.4.9.11. Shaper 91-99
- 2.4.9.12. Shaper 100-108
- 2.4.9.13. Shaper 109-117
- 2.4.9.14. Shaper 118-120
- 2.5. AIFB 1
- 2.6. AIFB 2
- 2.7. Measured Values
- 2.8. Filtering
- 3. Diagnostics
- 4. Service
- 5. Communication
- 6. Security
- 7. Language
- Appendix A: Alphabetical Parameter List
- Appendix B: Troubleshooting
- Appendix C: Maintenance
- Appendix D: Technical Reference
- Principles of Operation
- Echo Processing
- Measurement Range
- Measurement Response
- Damping
- Loss of Echo (LOE)
- Temperature derating curves
- Appendix E: Communications
- Appendix F: Firmware Revision History
- Glossary
- Index
- LCD menu structure
7ML19985LY01 SITRANS LR560 (FF) – INSTRUCTION MANUAL Page 135
mmmmm
B: Troubleshooting
Reading does not
change, but the level
does
SITRANS LR560 process-
ing wrong echo, i.e. vessel
wall, or structural member
• re-locate SITRANS LR560
• check nozzle for internal burrs or
welds
•use Auto False Echo Suppression
(2.4.8.1.) if necessary: see Auto False
Echo Suppression (2.4.8.1.) on page
100
Measurement is con-
sistently off by a con-
stant amount
• setting for Low
Calibration Point (2.3.5.1.)
not correct
• setting for Sensor Offset
(2.3.5.3.) not correct
• check distance from sensor
reference point to Low Calibration
Point (2.3.5.1.)
•check Sensor Offset (2.3.5.3.)
Screen blank
Power error • check nameplate rating against
voltage supply
• check power wiring or source
Reading erratic Echo confidence weak • refer to Confidence (2.4.7.1.)
•use Auto False Echo Suppression
(2.4.8.1.) and Auto False Echo
Suppression Range (2.4.8.2.)
• use foam deflector or stillpipe
Material filling • re-locate SITRANS LR560
Reading response
slow
Fill Rate/Min (2.3.6.2.) set-
ting is incorrect
• increase measurement response if
possible
Reads correctly but
occasionally reads
high when vessel is
not full
• detecting close range
echo
• build up near top of
vessel or nozzle
• nozzle problem
• use the air purge feature to clean the
antenna
•use Auto False Echo Suppression
(2.4.8.1.) and Auto False Echo
Suppression Range (2.4.8.2.)
Level reading lower
than material level
• material is within Near
Range zone
• multiple echoes
processed
• decrease Near Range (2.4.1.):
minimum value depends on antenna
type
• raise SITRANS LR560
• ensure Algorithm (2.4.5.1.) is set to tF
(trueFirst echo)
• vessel near empty and
low dK material
•set Position Detect (2.4.5.2.) to Hybrid
• check the setting for CLEF Range
(2.4.5.4.)
Operation Troubleshooting (cont’d)
Symptom Cause Action