7ML19985LY01 SITRANS LR560 (FF) – INSTRUCTION MANUAL Page 133
mmmmm
B: Troubleshooting
S: 31 Flash error. Repair required: contact your local
Siemens representative.
S: 33 Factory calibration for the internal tem-
perature sensor has been lost.
Repair required: contact your local
Siemens representative.
S: 34 Factory calibration for the device has
been lost.
Repair required: contact your local
Siemens representative.
S:39 Transducer temperature sensor failure. Repair required: contact your local
Siemens representative.
S:40 Transducer temperature too high. Relocate device and/or lower
process temperature enough to cool
device. Inspect for heat-related
damage and contact your local
Siemens representative if repair is
required.
S: 41 Transducer temperature too low. Relocate device and/or raise process
temperature enough to warm device.
Inspect for temperature-related
damage and contact your local
Siemens representative if repair is
required.
S:64 Device error.
NOTE: Fault text and icon appear only
on LCD.
Repair required. Contact your local
Siemens representative.
S:66 to
S:83
Device error.
NOTE: Fault text and icons appear only
on LCD.
Repair required. Contact your local
Siemens representative.
General Fault Codes (cont’d)
Code
/Icon
Meaning Corrective Action