General Fault Codes
2 Simulation or substitute value
Communication is good; device is in simulation mode or works with
substitute values.
3 Out of operation
Communication is good; device is out of action.
No data exchange
Write access enabled
Write access disabled
Notes:
If more than one fault is present, the device status indicator and text for each fault
alternate at 2 second intervals.
General Fault Codes
Code
/Icon
Meaning Corrective Action
S: 0 The device was unable to get a mea-
surement within the LOE Timer period.
Possible causes: faulty installation,
antenna material buildup, foaming/
other adverse process conditions,
invalid calibration range.
Ensure installation details are cor-
rect.
Ensure no material buildup. Clean if
necessary.
Adjust process conditions to mini-
mize foam or other adverse condi-
tions.
Correct range calibration.
If fault persists, contact your local
Siemens representative.
S: 2 Unable to collect profile because of a
power condition that is outside the
operating range of the device.
Repair required. Contact your local
Siemens representative.
Icon
Priority
Level
Meaning (cont’d)