User's Manual

Table Of Contents
Page 120 SITRANS LR560 (FF) – INSTRUCTION MANUAL 7ML19985LY01
mmmmm
Parameters
4.11. Secondary Value
The value displayed in the secondary region of the LCD, in Measurement Mode . [See
The LCD Display on page 31, area (6) under Normal operation.]
Use Secondary Value to capture the menu navigation path to a selected parameter,
and store a custom secondary value [for example, Echo Strength (2.4.7.2.)].
While in Parameter View
1)
mode of the selected parameter, press the decimal point
key. This stores the path to the selected parameter in Secondary Value, and displays
that value in the secondary region of the LCD display when in Measurement Mode.
4.12. Simulate Enable
Replaces a physical jumper switch found on some FF devices to enable simulation
when set to ON. (Available only via local operation.)
For more information on Simulation, see
Simulation (Input)
on page 51 in AMS
Device Manager, or the manual,
Foundation Fieldbus for Level instruments
(7ML19985MP01)
.
4.13. Demo Mode
Reduces the time between measurements and the accuracy for demonstration
purposes.
5. Communication
5.1. Tag
The user-defined description for the device.
To access this parameter via AMS Device Manager see Identification under
Identification (RESOURCE)
on page 61.
1)
See
Parameter view
on page 31, under
Program mode
.
Options
* OFF Simulation Disabled
ON Simulation Enabled
Note: Default settings in the parameter tables are indicated with an asterisk (*) unless
explicitly stated.