User's Manual

Table Of Contents
7ML19985KB01 SITRANS LR560 (mA/HART) – INSTRUCTION MANUAL Page 93
mmmmm
Parameters
2.7.5.3. Average Amount
The fraction of the old shot data that is kept for averaging purposes. A
higher value will give a smoother profile at the expense of a slower echo
profile response.
2.7.6. Echo Quality
2.7.6.1. Confidence
Indicates echo reliability: higher values represent better echo quality. The
display shows the echo confidence of the last measurement. Echo
Threshold (2.7.3.2.) defines the minimum criterion for echo confidence.
Open the menu Device – Echo Profile Utilities and click on the tab Echo
Profile.
2.7.6.2. Echo Strength
Displays the absolute strength (in dB above 1 μV rms) of the echo selected
as the measurement echo.
Open the menu Device – Echo Profile Utilities and click on the tab Echo
Profile.
2.8. TVT setup
2.8.1. Auto False Echo Suppression
Used together with Shaper Mode (2.8.4.) to screen out false echoes in a vessel
with known obstructions.
A ’learned TVT’ (time varying threshold) replaces the
default TVT over a specified range.
• See
Shaper Mode (2.8.4.)
on page 126 for a more detailed explanation.
To set via SIMATIC PDM, see
Auto False Echo Suppression
on page 46.
To set via local operation, see
AFES (Auto False Echo Suppression) Wizard
on page 77.
Values
0.0 to 1.0
Default: 0.75
Values (view only)
0 to 99
---- Shot not used
Related Parameters
Echo Threshold (2.7.3.2.)
Values (view only)
–20 to 99
Note: Default settings in the parameter tables are indicated with an asterisk (*)
unless explicitly stated.
Note: We recommend setting Auto False Echo Suppression either via
SIMATIC PDM or via local operation using the Auto False Echo Suppression
wizard.