User's Manual

Table Of Contents
7ML19985KB01 SITRANS LR560 (mA/HART) – INSTRUCTION MANUAL Page 57
mmmmm
AMS Device Manager
Quick Start Wizard via AMS Device Manager
A Wizard provides an easy step-by-step Quick Start procedure that configures the device for
a simple application.
Quick Start
Step 1 – Identification
1) Navigate to Configure/Setup > Wizard > Quick Start.
2) Click on Step 1.
3) You can accept the default values without modification. (Descriptor, Message, and
Date fields can be left blank.) If desired, make changes then click on Apply.
Step 2 – Application Type
1) Click on Step 2.
2) Select the application type (Steel or Concrete) and the operation (Level, Space, or
Distance
1)
) .
Notes:
The layout of the dialog boxes shown may vary according to the resolution setting
for your computer monitor
At each step, you can accept the default values without modification and click on the
next step to proceed.
After modifying parameters click on Apply inside the Quick Start window to write the
new values to the device.
Click on OK only if you wish to update all parameters to the device and close AMS.
Notes:
Selecting STEEL changes the setting for Position Detect (2.7.3.3.) to Rising Edge. and
for Algorithm (2.7.3.1.) to F.
Selecting CONCRETE changes the setting for Position Detect (2.7.3.3.) to Rising Edge
and for Algorithm (2.7.3.1.) to ALF.
1)
For details see
Sensor Mode (2.2.2.)
on page 129.