7ML19985KB01 SITRANS LR560 (mA/HART) – INSTRUCTION MANUAL Page 39
mmmmm
SIMATIC PDM
Quick Start
Launch SIMATIC PDM, open the menu Device – Wizard – Quick Start, and follow steps 1 to 4.
Step 1 – Identification
1) Click on Read Data from Device to upload Quick Start parameter settings from the
device to the PC/PG and ensure PDM is synchronized with the device.
2) Click on NEXT to accept the default values. (Description, Message, and Installation
Date fields can be left blank.)
Notes:
• The Quick Start wizard settings are inter-related and changes apply only after you
click on FINISH AND DOWNLOAD at the end of the last step to save settings offline
and transfer them to the device.
• Do not use the Quick Start Wizard to modify individual parameters: for quick access
to echo profile parameters, see
Echo profile
on page 44, or see
Parameter
Reference
on page 76 for a complete list. (Perform customization only after the
Quick Start has been completed.)
• Click on BACK to return and revise setting or Cancel to exit the Quick Start.
• For a vessel with obstructions, see
Auto False Echo Suppression
on page 46.
Note: The layout of the dialog boxes shown may vary according to the resolution
setting for your computer monitor.
Quick
Start