7ML19985KB01 SITRANS LR560 (mA/HART) – INSTRUCTION MANUAL Page 35
mmmmm
Quick Start: local
Requesting an Echo Profile
• In PROGRAM mode, navigate to: Level Meter >
Diagnostics (3.) > Echo Profile (3.2.).
• Press RIGHT arrow to request a profile.
•Use UP or DOWN arrow to scroll to an icon. When an icon is highlighted,
that feature becomes active.
• To move a cross-hair, press RIGHT arrow to increase the value, LEFT arrow
to decrease.
• To Zoom into an area, position the intersection of the cross-hairs at the center of
that area, select Zoom, and press RIGHT arrow . Press LEFT arrow to Zoom
out.
• To update the profile, select Measure and press RIGHT arrow .
• To return to the previous menu, select Exit then press RIGHT arrow .
algorithm: tF (trueFirst)
confidence
distance from flange face to target
TVT
echo
measure
exit
distance from Low Calibration Point to vertical cross-hair
zoom
pan left/right - selected
pan up/down
Echo amplitude (in dB)